CVE-2019-25719
8.6 HIGHDräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower con...
Published: 2026-06-02 · Last updated: 2026-06-03
Severity and scoring
- CVSS
- 8.6 HIGH
- Vector
- CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H
- CWE
- CWE-924
Description
Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.
Source: NVD
References
Related CVEs
Same CWE
- CVE-2026-39827 — An authenticated SSH client that repeatedly opened channels which were rejected by the server caused unbounded memory growth, eventually ... (6.5 MEDIUM)
- CVE-2023-2885 — Improper Enforcement of Message Integrity During Transmission in a Communication Channel vulnerability in CBOT Chatbot allows Adversary i... (8.1 HIGH)